Pooling Nasopharyngeal Swab Specimens to Improve Testing Capacity for SARS-CoV-2 by Real-Time RT-PCR

Abstract

Background

The detection of SARS-CoV-2 using qRT-PCR with the pooling of samples can reduce workload and costs especially when the prevalence rate of COVID-19 in a population is low.

To analyse the effect of pooling samples on the sensitivity of RT-qPCR for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection, we compared the cycle threshold (Ct) values of pools of 5 and 10 that tested positive with Ct values of individual samples that tested positive in that pool. Twenty positive nasopharyngeal (NP) specimens with low to high viral load were selected and pooled individually with four and nine negative NP.

Results

In NP specimens, the sensitivity of pools of 5 and 10 were 90 and 85%, compared to individual sample testing, respectively. The RT-qPCR sensitivity of pools of 5 and 10 against individual testing were not significantly different (p > 0.05). Detection of positive samples with low Ct values (< 36) was consistently achieved in pools of 5 and 10. However, there were higher false negatives when samples with high ct values (> 36) were pooled and tested. The mean Ct values obtained with the 5-sample pooled testing exceeded individual sample testing by 1.85 ± 1.09 cycles, while Ct values obtained with the 10-sample pooling exceeded individual sample testing by 3.4 ± 1.65 cycles.

Conclusions

In a low prevalence setting, testing capacity can be increased by pooling 5 or 10 samples, but the risk of additional false negatives needs to be considered.

Introduction

Coronavirus Disease 2019 (COVID-19) pandemic has caused the worldwide public health emergency, the most serious one for the past 100 years. The World Health Organization estimates on 9 October 2020 over 36 million infected individuals and close to 1,049,810 reported deaths worldwide (https://covid19.who.int/). After the highest peaks in Asiatic and European countries, come the turn of the American and the African continents. The African countries are hit hard by the Covid-19 pandemic for a number of reasons. One particular reason is the limited availability of sufficient reagents and PCR Kits for the diagnosis of SARS-CoV-2 infection. In time, COVID-19 testing is one of the most important tools used to contain the pandemic and break the chains of transmission. In developing countries like in Tunisia, the need to increase testing capacity and conserve reagents has become the main concern in testing. Testing specimens in pools has already been used in blood banks around the world to detect viruses that can transmit blood transfusions, such as human immunodeficiency virus (HIV), hepatitis B and C viruses, and even in investigating the influenza pandemic (H1N1) of 2009 [1,2,3]. In the United States, pooling was used retrospectively to screen COVID-19 in specimens in the early period of the pandemic when viral circulation was low and this is because of its simplicity and cost-effectiveness [4].

Pooled specimens before extraction result in a significant reduction in the number of tests performed and would be more economical than individually testing especially by asymptomatic individuals. Successful application of pooling specimens depends upon knowledge of the limit of detection, sensitivity, and specificity of the assay, and the prevalence rate of the disease. Therefore, our main goal was to evaluate the efficiency of specimen pooling for testing of SARS-CoV-2 virus using the commercial kit “genesig real-Time PCR COVID-19.

Materials and Methods

The study was approved by the ethics committee and medical research of the Farhat Hached University Hospital, Sousse, Tunisia (Reference number IRB00008931). This study was an evaluation of laboratory techniques using conserved clinical specimens.

Sample Pooling

Nasopharyngeal (NP) specimens used in this study had been collected as part of clinical care from patients under investigation for COVID-19 infection placed in 2 mL of universal viral transport medium and sent to the laboratory of Microbiology of the University Hospital of Farhat Hached of Sousse. A total of 50 specimens had been tested negative for SARS-Cov2 using the available emergency authorized qualitative real-time PCR assays. A total of 20 SARS-CoV-2-positive samples were used for pooling. The selected Ct value of individual SARS-CoV-2 positive samples was spanning a large range of Ct of less than 40. Specimens with Ct values between 26 and 36 were considered to have low RNA viral load, while those with Ct values lower than 26 were considered to have a high viral load. Specimens with Ct values higher than 36 were considered weakly positive. All specimens had been stored at − 80 °C. Each positive nasopharyngeal swab specimen (0.140 mL) was mixed with 4 and 9 negative nasopharyngeal swab samples (0.140 mL each), obtaining pools of final volume 0.7 mL and 1.4 mL respectively. Out of 40 mini-pools either pools of 5 (n = 20) or 10 (n = 20) samples were made, with each containing one of the positive NP specimens (140 mL) mixed with 4 or 9 negative specimens. These pools were then vortexed for 5–10 s before RNA extraction and PCR testing.

RNA Extraction and Real-Time PCR Assay

RNA was extracted from 0.140 mL of each individual and pooled specimen using the QIAamp viral RNA mini kit in conjunction with a QIAcube (QIAGEN, Hilden, Germany) following the manufacturer’s instructions. PCR amplification was performed using the Primerdesign Ltd. COVID-19 genesig assay targeting the (ORF1ab) genome region on the Applied Biosystem 7500 Real-Time PCR System (Applied Biosystems, Foster City, CA, USA) according to the manufacturers’ instruction. PCR was performed in duplicate (replicates I and II) for each extracted specimen. RNA internal extraction control was used to identify PCR inhibition and to evaluate RNA extraction purity and integrity of the PCR run. The detection limit of ORF1ab was ≤0.33 copies/μL. To improve the sensitivity of the pooling method, the number of cycles has been increased to 50. According to the laboratory’s protocol, samples with Ct > 36 that tested weakly positive with only one of the replicates are defined as inconclusive and required confirmation by retesting.

Statistical Analysis

Statistical analysis was performed using SPSS software (version 22.0). The cycle thresholds (Ct) for the amplification of the ORF1ab gene were analyzed. The results are expressed as the average (mean) ± standard deviation (SD). Statistical significance of mean Ct differences (ΔCt) among pools of 5 and 10 were detected by Student’s T-test. P < 0.05 was considered statistically significant.

Results

Comparison of Ct between the original and pooled SARS-CoV-2 positive samples are detailed in Table 1. PCR inhibition was not observed in all the NP specimens used in the present study and minimal variations were observed in all the internal control CT values (data not shown). Positive pools of 5 and 10 samples could still be confidently identified, even if Ct values of single samples were up to 36. In pooled testing of 5 samples (Ct > 36), only two low positive specimens have been determined inconclusive and when tested in a pool of 10 samples, one positive specimen has been determined to be inconclusive, and two positive specimens have been determined to be undetected for both replicates. The overall sensitivity of the pool testing approach, with 5 and 10 specimens per pool, were 90 and 85% compared to individual tests. The sensitivity, of RT-qPCR pool-5 and pool-10 against individual testing, were not significantly different (p > 0.05). Our results showed that mean Ct values of the ORF1ab gene in both positive pools of 5 and 10 were 33.56 ± 4.1 and 34.9 ± 4.34 respectively, significantly higher than that of the un-pooled samples (p = 0.03, p = 0.001, respectively), indicating reduced viral loads. For these samples, the mean Ct values differences (ΔCt) between individual positive samples and pooled tests were 1.85 ± 1.09 in pools of 5, and 3.4 ± 1.65 in pools of 10. As expected, the mean Ct differences (ΔCt) between the original and each dilution aligned well with results expected for the given dilutions (i.e., In a real-time RT-PCR assay with 100% efficiency a 1:10 dilution leads to a 3.3 Ct value rise).

Table 1 Detection of SARS-CoV-19 RNA by RT-PCR in pooled nasopharyngeal (NP) specimens from patients with COVID-19

Discussion

Due to the huge number of tests that are being carried out universally, reagents needed for the SARS-CoV-2 testing are in short supply to ensure the rapid diagnostic process in patients with COVID-19 [5, 6]. Pooling specimens have been shown to increase testing capacity and reduce the cost per test for the detection of viral infections such as HIV, hepatitis B and C virus [1, 3]. A key step to control the epidemics is to implement fast and sensitive approaches for diagnostic. In the current study, pooling specimens were used to detect SARS-CoV-2 in NP specimens using the Primerdesign Ltd. COVID-19 genesig assay. Our data suggest that pooling of 5 or 10 samples per pool can increase test capacity with the existing equipment and test kits and detects positive samples with sufficient diagnostic accuracy. In contrast to developed countries, detailed published data on the practical use of pooled testing in outbreaks in resource-limited settings are lacking. There is only scarce information available in Tunisia [7]. The main limitations of the literature published so far are the small number of cases processed by this technique [8], mathematical models and the theoretical considerations [9, 10]. Three different approaches of pooling have been proposed: to pool swab samples from different patients during the collection process into a single volume of transport media (VTM), to pool the VTM of the samples from different samples to create a homogeneous pool in the laboratory and to pool the nucleic acid where an aliquot of RNA extracted from each sample is collected to create a homogeneous pool [11]. For laboratories, considering the positivity rate should be taken into consideration before implementing a pooling approach because pool testing is most efficient when applied in settings of low virus prevalence which may be more likely in an asymptomatic population especially in settings with limited capacity such in African countries. On the basis of our laboratory results, large cohorts and, testing on asymptomatic individuals can be carried out in particular in limited settings. The same as with previous studies when the incidence rate of SARS-CoV-2 infection is 10% or less, pool testing will result in the saving of reagents and personnel time with an increase in the testing capability of at least 69% [7, 12]. Pooling is not efficient during periods of high prevalence rate (>15%) when every pool yields a positive result and thus demand subsequent retesting to every specimen in the pool separately [1]. One strategy to avoid this is to adjust the size of the pool according to the infection prevalence. It is recommended that laboratories evaluate the pooling approach against the disease prevalence in the geographical area they serve before using this for routine diagnostic purposes.

The optimal number of pooled samples depends on the laboratory’s protocols and especially the limit of detection of the RT-PCR assay. Sensitivity for the various pool sizes can be improved with repeated sampling. The main disadvantage of pool testing is that there might be a decrease in test sensitivity. The sensitivity of the pooling is affected by many factors such as the sensitivity of the kit, the used dilution, collection sample techniques, samples type (NP, oropharyngeal, nasal, etc.), sample transport temperature, and viral load [13,14,15].

A recent study showed very encouraging results for SARS-CoV-2 using pools of up to 7 samples before the extraction and up to 60 samples after the RNA was extracted with false negativity of 10% [16], but this will mainly depend on validation protocol. Similar to our study, Wacharapluesadee et al. [17] used a pool size of 10 and found a 13.3% of false-negative due to positive samples with a low viral load. One of the major factors affecting the test sensitivity is the dilution effect of the pooled samples, leading to a higher rate of false negatives [18, 19]. The similar pooling strategy approach was presented by Sawicki, R et al. [8] who randomly pooled one positive sample and five negative individual NP swabs and one positive sample and eight negative individual samples, and who recommended concentration before the extraction step. The study results showed that pool testing could detect even up to a single positive sample with a Ct value as high as 34. In the current study, the detection of a specimen with a Ct value < 36 was not compromised when pooled with four or nine negative specimens. However, weak positive individual samples (Ct value > 36) might escape detection in pools. This is in line with recent works that reported a decreased sensitivity of pool testing at similar dilutions of pools of 5 and 10 in the presence of low viral load samples (Ct value > 36) [17,18,19,20,21]. Analysis of pool results requires close attention to inconclusive-result pools, as these may contain individual positive samples [1]. It is therefore advisable to test again a patient on whom there is a high clinical suspicion of COVID-19 infection even if the first test was negative. Weak positive samples were seen typically in convalescent patients after the resolution of symptoms [22]. These patients were not likely to be infectious and therefore were unlikely to be able to transmit the virus to another person [23,24,25,26]. In another study, no dilution errors were reported in very small pools (n = 5) [12]. Moreover, Yelin et al. [16] have concluded that the dilution effect is minimal for groups of 32 samples, but this is not a consistent finding in all studies. Similar to our study pools consisting of many samples may require additional amplification cycles due to lower viral load in the pooled samples [16, 27].

The dilution effect due to pooling was considered to be of varying importance among the studies, which is likely because, in the different experimental settings, samples with different viral loads were pooled. In this study, we found that the probability of detection of the positive pool decreased as the Ct value of the single positive sample in the pool increased (lower viral load). The mean Ct differences of the ORF1ab target of specimens when originally tested compared to undiluted specimens after one freeze-thaw was very small. On average, Ct values obtained with the 5-sample pooled testing exceeded individual sample testing by 1.85 ± 1.09 cycles while Ct values obtained with the 10-sample pooling exceeded individual sample testing by 3.4 ± 1.65 cycles. The viral load has been suggested as another important factor affecting the sensitivity of the assay in pooled samples. Further, test sensitivity of pooling depends on the distribution of viral loads across individuals and over the course of infection (growth versus decay) with implications for appropriate surveillance and the interpretation of a viral load from a sample pool [28, 29]. Despite the fast growth of the literature on theoretically optimal pooling designs for COVID-19 testing, formal inclusion of biological variation (i.e., viral loads) and incorporation of general position along the epidemic curve has received minimal attention [28,29,30]. Understanding the accurate performance of pooled testing requires modeling the prevalence in prolonged or multiple epidemic periods and the dynamics of viral load [29].

The results of our laboratory analyses should be interpreted with caution because conducted on a limited number of specimens. Additional studies need to be performed in order to determine if a larger input volume than that recommended for the QIAamp viral RNA mini kit will provide better results using this pooling protocol. The procedure evaluated herein was improved by increasing the number of RT-PCR cycles. However, it is possible that this approach can be further improved by using fresh specimens and by collecting samples in a smaller volume of transport medium, or by collecting samples directly in a small volume of lysis buffer provided that specimens are processed without delay [20].

Due to the global shortage of reagents and kits for SARS-CoV-2 (CoV-2) tests, the implementation of a single marker assay rises as a feasible and cost-effective option for diagnostics especially in low and middle-income countries. However, recent studies suggest that SARS-CoV-2 assay targeting two positions in the genome may improve the efficiency of the detection of positive specimens in low-size mini pools [20].

Conclusion

According to the findings of this study, pooling samples for SARS-CoV-2 detection by real-time RT-PCR may be an acceptable method with minimal loss of sensitivity even for low virus loads. Pooling specimens will help to increase coronavirus testing capacity in order to meet the high demand for testing in the mass screening programs needed in the early identification and isolation of asymptomatic individuals especially during the constrained supply of reagents and PCR kits for the diagnosis of SARS-CoV-2 infection.

Availability of Data and Materials

The authors confirm that the data supporting the findings of this study are available within the article.

Abbreviations

COVID-19:

Coronavirus Disease 2019

Ct:

Cycle threshold

NP:

Nasopharyngeal

RT-qPCR:

Reverse transcription real time PCR

RNA:

Ribonucleic acid

SARS-CoV-2:

Severe acute respiratory syndrome coronavirus 2

VTM:

Volume of Transport Media

References

  1. Van TT, Miller J, Warshauer DM, Reisdorf E, Jernigan D, Humes R, et al. Pooling nasopharyngeal/throat swab specimens to increase testing capacity for influenza viruses by PCR. J Clin Microbiol. 2012;50:891–6.

    Article  Google Scholar 

  2. Emmanuel JC, Bassett MT, Smith HJ, Jacobs JA. Pooling of sera for human immunodeficiency virus (HIV) testing: an economical method for use in developing countries. J Clin Pathol. 1988;41:582–5.

    Article  CAS  Google Scholar 

  3. Mine H, Emura H, Miyamoto M, Tomono T, Minegishi K, Murokawa H, et al. High throughput screening of 16 million serologically negative blood donors for hepatitis B virus, hepatitis C virus and human immunodeficiency virus type-1 by nucleic acid amplification testing with specific and sensitive multiplex reagent in Japan. J Virol Methods. 2003;112:145–51.

    Article  CAS  Google Scholar 

  4. Hogan CA, Sahoo MK, Pinsky BA. Sample pooling as a strategy to detect community transmission of SARS-CoV-2. JAMA. 2020;323:1967–9.

    Article  CAS  Google Scholar 

  5. Babiker A, Myers CW, Hill CE, Guarner J. SARS-CoV-2 testing. Am J Clin Pathol. 2020;153:706–8.

    Article  CAS  Google Scholar 

  6. Esbin MN, Whitney ON, Chong S, Maurer A, Darzacq X, Tjian R. Overcoming the bottleneck to widespread testing: a rapid review of nucleic acid testing approaches for COVID-19 detection. RNA. 2020;26:771–83.

    Article  CAS  Google Scholar 

  7. Abid S, Ferjani S, El Moussi A, Ferjani A, Nasr M, Landolsi I, et al. Assessment of sample pooling for SARS-CoV-2 molecular testing for screening of asymptomatic persons in Tunisia. Diagn Microbiol Infect Dis. 2020;98:115125.

    Article  CAS  Google Scholar 

  8. Sawicki R, Korona-Glowniak I, Boguszewska A, Stec A, Polz-Dacewicz M. Sample pooling as a strategy for community monitoring for SARS-CoV-2. Sci Rep. 2021;11:3122.

    Article  CAS  Google Scholar 

  9. Aragon-Caqueo D, Fernandez-Salinas J, Laroze D. Optimization of group size in pool testing strategy for SARS-CoV-2: a simple mathematical model. J Med Virol. 2020;92:1988–94.

    Article  CAS  Google Scholar 

  10. Brynildsrud O. COVID-19 prevalence estimation by random sampling in population - optimal sample pooling under varying assumptions about true prevalence. BMC Med Res Methodol. 2020;20:196.

    Article  CAS  Google Scholar 

  11. Millioni R, Mortarino C. Test groups, not individuals: a review of the pooling approaches for SARS-CoV-2 diagnosis. Diagnostics (Basel). 2021;11:68.

    Article  CAS  Google Scholar 

  12. Abdalhamid B, Bilder CR, McCutchen EL, Hinrichs SH, Koepsell SA, Iwen PC. Assessment of specimen pooling to conserve SARS CoV-2 testing resources. Am J Clin Pathol. 2020;153:715–8.

    Article  CAS  Google Scholar 

  13. Mahmoud SA, Ibrahim E, Thakre B, Teddy JG, Raheja P, Ganesan S, et al. Evaluation of pooling of samples for testing SARS-CoV- 2 for mass screening of COVID-19. BMC Infect Dis. 2021;21:360.

    Article  CAS  Google Scholar 

  14. Arevalo-Rodriguez I, Buitrago-Garcia D, Simancas-Racines D, Zambrano-Achig P, Del Campo R, Ciapponi A, et al. False-negative results of initial RT-PCR assays for COVID-19: a systematic review. PLoS One. 2020;15:e0242958.

    Article  CAS  Google Scholar 

  15. Afzal A. Molecular diagnostic technologies for COVID-19: limitations and challenges. J Adv Res. 2020;26:149–59.

    Article  CAS  Google Scholar 

  16. Yelin I, Aharony N, Shaer Tamar E, Argoetti A, Messer E, Berenbaum D, et al. Evaluation of COVID-19 RT-qPCR test in multi-sample pools. Clin Infect Dis. 2020;71:2073–8.

  17. Wacharapluesadee S, Kaewpom T, Ampoot W, Ghai S, Khamhang W, Worachotsueptrakun K, et al. Evaluating the efficiency of specimen pooling for PCR-based detection of COVID-19. J Med Virol. 2020;92:2193–9.

    Article  CAS  Google Scholar 

  18. Bateman AC, Mueller S, Guenther K, Shult P. Assessing the dilution effect of specimen pooling on the sensitivity of SARS-CoV-2 PCR tests. J Med Virol. 2021;93:1568–72.

    Article  CAS  Google Scholar 

  19. Baccini M, Rocco E, Paganini I, Mattei A, Sani C, Vannucci G, et al. Pool testing on random and natural clusters of individuals: optimisation of SARS-CoV-2 surveillance in the presence of low viral load samples. PLoS One. 2021;16:e0251589.

    Article  CAS  Google Scholar 

  20. Torres I, Albert E, Navarro D. Pooling of nasopharyngeal swab specimens for SARS-CoV-2 detection by RT-PCR. J Med Virol. 2020;92:2306–7.

  21. Praharaj I, Jain A, Singh M, Balakrishnan A, Dhodapkar R, Borkakoty B, et al. Pooled testing for COVID-19 diagnosis by real-time RT-PCR: a multi-site comparative evaluation of 5- & 10-sample pooling. Indian J Med Res. 2020;152:88–94.

    Article  CAS  Google Scholar 

  22. Lohse S, Pfuhl T, Berko-Gottel B, Rissland J, Geissler T, Gartner B, et al. Pooling of samples for testing for SARS-CoV-2 in asymptomatic people. Lancet Infect Dis. 2020;20:1231–2.

  23. Colavita F, Lapa D, Carletti F, Lalle E, Bordi L, Marsella P, et al. SARS-CoV-2 isolation from ocular secretions of a patient with COVID-19 in Italy with prolonged viral RNA detection. Ann Intern Med. 2020;173:242–3.

    Article  Google Scholar 

  24. Cheung KS, Hung IFN, Chan PPY, Lung KC, Tso E, Liu R, et al. Gastrointestinal manifestations of SARS-CoV-2 infection and virus load in fecal samples from a Hong Kong cohort: systematic review and meta-analysis. Gastroenterology. 2020;159:81–95.

    Article  CAS  Google Scholar 

  25. Zheng S, Fan J, Yu F, Feng B, Lou B, Zou Q, et al. Viral load dynamics and disease severity in patients infected with SARS-CoV-2 in Zhejiang province, China, January-march 2020: retrospective cohort study. BMJ. 2020;369:m1443.

    Article  Google Scholar 

  26. Li D, Jin M, Bao P, Zhao W, Zhang S. Clinical characteristics and results of semen tests among men with coronavirus disease 2019. JAMA Netw Open. 2020;3:e208292.

    Article  Google Scholar 

  27. Taufer M. Rapid, large-scale, and effective detection of COVID-19 via non-adaptive testing. J Theor Biol. 2020;506:110450.

    Article  CAS  Google Scholar 

  28. Cleary B, Hay JA, Blumenstiel B, Harden M, Cipicchio M, Bezney J, et al. Using viral load and epidemic dynamics to optimize pooled testing in resource-constrained settings. Sci Transl Med. 2021;13:eabf1568.

  29. Hay JA, Kennedy-Shaffer L, Kanjilal S, Lennon NJ, Gabriel SB, Lipsitch M, et al. Estimating epidemiologic dynamics from cross-sectional viral load distributions. Science. 2021;373:eabh0635.

  30. Cevik M, Tate M, Lloyd O, Maraolo AE, Schafers J, Ho A. SARS-CoV-2, SARS-CoV, and MERS-CoV viral load dynamics, duration of viral shedding, and infectiousness: a systematic review and meta-analysis. Lancet Microbe. 2021;2:e13–22.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

No funding was received for conducting this study.

Author information

Authors and Affiliations

Authors

Contributions

IH, NH, NBA and JB contributed to the conception and design of research. IH, MM, OH acquired data in the study. IH performed the experiments. IH and NH participated in the analysis and interpretation of data. IH and NH wrote the draft manuscript. NH, NBA and JB supervised and supported the study. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Imene Handous.

Ethics declarations

Consent to Publication

Not applicable.

Ethics Approval and Consent to Participate

The study was approved by the ethics committee and medical research of the Farhat Hached University Hospital, Sousse, Tunisia (Reference number IRB00008931).

Competing Interests

The authors have no conflicts of interest to declare.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Handous, I., Hannachi, N., Marzouk, M. et al. Pooling Nasopharyngeal Swab Specimens to Improve Testing Capacity for SARS-CoV-2 by Real-Time RT-PCR. Biol Proced Online 23, 19 (2021). https://doi.org/10.1186/s12575-021-00156-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12575-021-00156-6

Keywords