From: Cell-Wide Survey of Amide-Bonded Lysine Modifications by Using Deacetylase CobB
▲CRITICAL Prior to GC-MS analysis, ensure that the instrument has been optimized. Details for our GC/MS workflow are listed below. Please be aware that other GC/MS methods with sufficient sensitivity, reproducibility, and linearity for the metabolites of interest can be used. | |
1. Connect an Agilent 5977B Mass Spectrometer to an Agilent 7890B gas chromatograph. Use 5977MS/Enhanced MassHunter software to control the GC/MS system. | |
GC settings: | |
Column: Agilent 19091S-433HP-5MS 5% Phenyl Methyl Silox 325 °C: 29.8 m × 250 μm × 0.25 μm | |
Carrier gas: helium (> 99.999% purity) | |
Carrier gas flow rate: 1 mL/min | |
Oven temperature program: 100 °C for 3 min, increase the temperature by 10 °C/min to 140 °C, increase the temperature by 8 °C/min to 260 °C, increase the temperature by 10 °C/min to 310 °C, and hold at this temperature for 5 min. | |
Injection port temperature: 280 °C | |
Transfer line temperature: 250 °C | |
Injection volume: 1 μL | |
Mode: splitless | |
MS settings: | |
Ionization mode: electron ionization (EI) mode | |
Electron energy: 70 eV | |
Ion source temperature: 250 °C | |
Scan frequency: 2.7/s | |
Mass range: 50.00–600.00 m/z | |
2. Place the samples in the auto sampler tray for GC/MS analysis. | |
3. Create a sample list (or sequence list) that includes the metabolomic sample details. | |
4. Inject the pyridine standard at the beginning and the end of the sample analysis sequence. |